Concentration Meter

Simple wagon

  • Overview
  • Specifications

Overview

Reasonably priced wagon for a particular chemical solution. Ideal model for off-line control of chemical concentrations in various processes.

List of chemical solutions by device type

Features

Allows installation anywhere, with superior resistance to the environment

Usable in acidic or alcohol environments.

Reasonable price due to a simple structure

Achieves a reasonable price by limiting applicable solutions to SC-1, SC-2, and SPM.

Use example

Specifications

Measurement method Transmittance Spectroscopy measurement based on near infrared rays (0.8 - 1.4μm)
Measurement wavelength Eight wavelength spectroscopy based on optical filter
Measurement cell Flow Cell (3ml), sapphire for hydrofluoric acid, quartz for others
Data processing method Computer processing by KURABO’s unique principal component recurrence method
Number of applicable components 1 - 2
Minimum measuring time 20seconds or less (only for measurement, excluding suction and cooling time, etc.)
Ambient temperature 25ºC ± 5ºC
Setting temperature for measurement 25ºC ± 5ºC
Dimensions / Weight 250W x 330D x 282H (mm)
15kg (spectroscope unit)
240W x 300D x 150H (mm)
6kg (Data processing unit)
* Excluding projections in dimensions
Power supply DC24V ± 10% MAX 220W
Tube diameter IN: Ø3 x Ø2 OUT: Ø4 x Ø3 (mm) Joint fitting
Zero calibration Once a day or more, using pure water of 25ºC ± 1ºC
Concentration display Four digit display (two decimal places)